Please cite as: CSH Protocols; 2008; doi:10.1101/pdb.prot4994
| Protocol |
This protocol was adapted from "Desorption Electrospray Ionization: Proteomics Studies by a Method that Bridges ESI and MALDI," Chapter 6, in Proteomics: Methods Express (eds. OConnor and Hames). Scion Publishing Ltd., Oxfordshire, UK, 2007.
INTRODUCTION
Desorption electrospray ionization (DESI) allows in situ analysis of biological tissues. The analysis of less abundant protein constituents within a tissue sample often requires the removal of lipid species prior to analysis, similar to the situation with matrix-assisted laser desorption/ionization (MALDI). After removal of lipid constituents, the tissue can be treated with protease to degrade proteins present in the tissue. The tryptic products can be investigated directly from the tissue using DESI. The spectra obtained feature ions of tryptic fragments from abundant proteins present in the tissue sample. The digestion is usually not complete; hence, the presence of missed cleavage sites is typical in the peptides detected. The signal is more stable for longer times than in the case of deposited samples, so the recording of mass spectrometry (MS)/MS data is simple in this case. DESI-MS is an emerging technique with great promise, but its application range is still being investigated. Therefore, the protocol for DESI-MS analysis of tissue sections presented here provides general procedures used for the applications that have been investigated so far. Optimal ion source parameters and surface types may vary depending on the application.
RELATED INFORMATION
An introduction to DESI instrumentation, methods, and applications is provided in the CSH Protocols article Desorption Electrospray Ionization: Proteomics Studies by a Method that Bridges ESI and MALDI (this issue). Protocols for Desorption Electrospray Ionization (DESI) Analysis of Intact Proteins/Oligopeptides and Desorption Electrospray Ionization (DESI) Analysis of Tryptic Digests/Peptides are also available (this issue).
Wiseman et al. (2005) analyzed adenocarcinoma tissue from human liver and determined the lipid composition at a series of positions across the tissue. In these experiments, thin frozen tissue sections were analyzed directly in the laboratory environment without additional treatment. The results from these studies indicated that the abundance of specific lipid species was enhanced in the tumor portion of the tissue. In situ analysis of unprocessed tissue is typified by the DESI mass spectrum shown in Figure 1A . In this case, an intact Gallus gallus (chicken) heart was analyzed using DESI. The most abundant ions present in the mass spectrum were phospholipids and other lipid species. Figure 1B shows the DESI mass spectrum recorded on a dried blood spot on the G. gallus heart. The analysis of the blood residue on the chicken heart showed the presence of heme and hemoglobin chains, as would be expected.
![]() View larger version (30K): [in a new window] |
Figure 1. In situ DESI analysis. (A) In situ DESI analysis of intact chicken (G. gallus) heart using ethanol/H2O spray solvent at 5 µL/min. (B) DESI mass spectrum of hemoglobin (Hgb) chains recorded on a dried blood spot on the chicken heart. (m/z) Mass-to-charge ratio. (Reprinted with permission, © 2007 Scion Publishing Ltd.)
|
Reagents
Spray solvent (e.g., methanol/H2O or aqueous buffers)
Typical spray solvents are acidified aqueous solvents, for example, 0.1 % acetic acid in H2O.
Tissue sections (fresh or frozen)
Equipment
Mass spectrometer equipped with DESI ion source
Pipette tips
Spray chamber
Surface slides (polymethyl methacrylate [PMMA], polytetrafluoroethylene [PTFE], or glass)
Vacuum desiccator
METHOD
| ||||||||||||||||||||
See Troubleshooting.
TROUBLESHOOTING
Problem: There are no surface-originated ions in the spectra.
[Step 6]
Solution: A possible cause of this is that the sprayed droplets and ions are not reaching the surface because of incorrect spray parameters (i.e., the volumetric flow rate is too low or there is no applied high voltage) or surface charging effects. Always check the spray pattern on the surface. If the spray is not visible on a glass surface, check to ensure that the solvent syringe pump is on and that there are no blockages in the solvent delivery line that would restrict flow.
Problem: No signal is detected.
[Step 6]
Solution: There may be no spray present, or the sprayed species are being deflected by the charged surface.
Try the following:
Problem: There is excessive adduct formation.
[Step 6]
Solution: This may be caused by insufficient desolvation due to the presence of large droplets or contamination of the surface. Increase the gas flow rate and/or heat the surface.
Problem: A transient signal, not suitable for MS/MS, is observed.
[Step 6]
Solution: A possible cause of this is lack of sample adhesion to the surface. The following are possible solutions:
Problem: Results exhibit strong suppression effects and poor spectral resolution.
[Step 6]
Solution: This may be caused by an inappropriate solvent system. Change the solvent composition.
REFERENCES
Wiseman, J.M., Puolitaival, S.M., Takats, Z., Cooks, R.G., and Caprioli, R.M. 2005. Mass spectrometric profiling of intact biological tissue by using desorption electrospray ionization. Angew. Chem. Int. Ed. Engl. 44: 7094–7097.
Copyright © 2008 by Cold Spring Harbor Laboratory Press. Online ISSN: 1559-6095 Terms of Service |